PhysioLab Software Guide
Contents
2011
MBP-SG102204
Installation
and
Update Instructions. 2
Configuring
the
Software. 5
Choosing a
Color
Scheme. 8
Overview
of the
Software Interface. 9
Recording
Data. 17
Event
Marking. 17
Setting Up
Tasks. 17
Managing
Data. 23
Adding and
Deleting Client Information.. 28
Display
Types. 34
EEG
Displays. 41
Recording
EEG
Data. 47
EMG
Displays. 49
ECG,
HR, and HRV Displays. 53
Respiration
Displays. 54
Breathing
Pattern. 55
Skin
Conductance and Skin Resistance Displays. 56
Temperature
Displays. 56
Audio
Features. 58
Setting Up
Windows
Media Player Feedback. 61
Installation and Update Instructions
System Requirements
The
operating system requirement is Windows 98 Second Edition or later with
all
Microsoft updates and Media Player 9 installed (WindowsXP recommended). The minimum hardware requirements are a
processor speed of 1 Gigahertz, 512 MB of RAM and a video card with 64
MB
RAM. In order to use the optional dual
monitor mode, you will need to run the software on either a laptop
computer or
on a desktop computer with a dual head video card.
Previously Installed Version Instructions
If you have a previous
version of USE3 Physiolab software installed, you do not need to
uninstall it
before performing a new installation.
Your client data will be safely retained. NOTE:
Any customized settings, such as saved
settings files and protocols, will be lost whenever you perform a new
install
or update the software with a download from the J&J Engineering
website. You can save customized files
by going to the C:\Program Files\J&J Engineering\Physiolab folder
and
renaming the .mdb file for each application you wish to save. These renamed customized applications will
be retained in their current version and not updated.
Return to Index
Installing Windows USE3 Physiolab
Software from
CD-ROM
Click Start, then My
Computer. Then double-click on the
letter of your CD-ROM
drive. Double click the blue Setup icon.
You will then get a series of
screens.
Welcome screen - click
Next.
Software License Agreement –
click Yes
User Information – type in
your name and company and then click Next.
Destination Location –
defaults to C:\Program Files\Physiolab\ Physiolab. Please accept the
default
location because future upgrades will automatically install to this
location. Click Next
Accept the default Program
Folder, click Next
Files are then copied.
When setup is complete,
click Finish
Click the X to close
the window.
Downloading Updates From Our Website
The update at jjengineering.com
on the Tech
Support page replaces only some of the files of your original CD-ROM
installation – please do not uninstall the software before downloading
an
update.
If you already have the
software installed and have created customized settings and
tasks, you
will have to re-create these after upgrading.
If you wish to save customized settings and protocols, you must
rename
the application file name. (For instructions, please see “NOTE” near
the top of
this page.)
To
download the upgrade:
On
the Tech Support webpage, click the heading called “Download (date)
USE3
Physiolab Upgrade”. Click "Open".
Click
“Next“ when prompted. Accept all
default settings (including the “Repair” selection).
Click
“Finish”.
If
you want to save the update to removable media for installation
on a
computer not connected to the internet, click “Save” instead of
“Open”. In the drop down box, browse to
your preferred save location. Accept the default file name and click “Save”.
Contacting J+J Engineering
Technical Support
Please visit J+J Engineering on the
World Wide Web at
www.jjengineering.com. Here
you can see the latest equipment,
download software upgrades and view technical support information.
Please
feel free to contact J&J Engineering whenever you feel that you
need
help. Also, if you would like to see
something different in a program or have product development ideas, we
would
appreciate your suggestions.
Phone:
1-360-779-3853
Fax:
1-360-697-4435
Email:
engineer@jjengineering.com
Send letters to J+J Engineering at:
J+J Engineering, Inc.
22797 Holgar Ct. NE
Poulsbo, WA98370
Return to Index
Configuring the Software
Setting Up the Software to Detect Your Hardware
Model
You will need to follow the
instructions below after your initial installation and after each
software
upgrade.
If Physiolab is not already
running, click on the Physiolab desktop icon
to
get to the opening window. Click Run Session to bring up the Select
Application window.
IMPORTANT:
Click on the drop-down
arrow in the Hardware
Type box and select the type of
J&J device
that you have connected. (The USE3
software supports many different hardware devices so it is important
that you
have selected the specific hardware that you are using.)
In the application list, click on the
name of the application you plan to run.
Next click Setup.

In the Hardware Setup window,
be sure that the correct port (COM or USB)
that you used to connect your device is selected. If
you are using a COM port connection, be sure the COM port number
matches the one you are using. NOTE:
This window displays the application date and currently installed
version date
(MBP Version)
THE LATEST VERSION IS 7-11-2007. There was a USB update in 2009, but it
did not include Physiolab.
Eliminating Electrical Line Frequency Noise
In North America the
electrical line frequency is 60Hz, but in many countries line frequency
is 50
Hz. The notch filter is defaulted to 60Hz so, if you live in a country
where
the electrical line frequency is 50 Hz, you will need to configure the
software
for your area. In the Hardware Setup
window (pictured above) check to see that the line frequency for your
area is
selected in the Notch filter box. If the
setting is incorrect you will see noise appearing as spikes at regular
frequencies in the signal FFT displays. (For additional information
please see
the “Artifact Detection” section in your Hardware Manual.)
Setting Up Dual Screen Mode
Many laptop computers support a
second monitor with two different simultaneous displays.
Many desktop computers can be upgraded to
this capability with the addition of a dual-monitor video card
available at
most computer stores.
The optional Dual Screen
Mode feature allows you to have a Master Screen with all the controls
for the trainer and a separate Dual Screen
with just a simple
feedback
display or game for the client.
Return to Index
Set-Up and Testing of Dual
Monitor Displays
If you have this capability
on your computer, it can be turned on as follows:
Right-click on the desktop,
then click Properties, then Settings.
Right-click on “2”, then click Attached. (A check
mark should appear next to “Attached”.) Check the box labeled Extend
my
Windows Desktop. Set the screen
resolution to the same resolution as primary screen #1, preferably
1024x768. Please see your computer
owner’s manual for help with this set-up.
When this is set up
properly, your computer should boot-up with your normal desktop on
Monitor #1
and the desktop background picture without any icons on Monitor #2. You should be able to run USE3 software with
dual displays. If you have trouble
getting the dual monitor desktop function working, please check with
your
computer supplier.
Using Dual Display Mode with USE3 Physiolab
Software
The USE3 software supports
dual monitors on specific screens that are marked in the screen name
with
letter D or Dual. To enable this
feature in USE3, from the USE-3 Sign-On Screen click on Options
and check mark Dual. Do
not do this unless you are sure dual
monitor mode with extended desktop is working on your computer.
Click the Dual Display
Icon
to
see a list of all available Dual Screen Displays
in an application, To select a feedback screen to appear on a second
monitor,
click on a dual screen display name and click APPLY.
The second monitor display
will have a feedback only with no controls or mouse functions. All the settings on the feedback screen will
be slaved to a display on Monitor #1.
(The exception to this is Windows Media Player used as a
feedback
screen.)
In Dual Mode on laptop
computers you may notice a speed reduction when the screen is updating. This is due to your computer’s display
driver limitations. It should not
affect feedback.
In most applications the
feedback display on Monitor #2 will also be shown on Monitor #1 as a
smaller
sub-display. To make changes in the
feedback display, you need to highlight the display and signal on
Monitor #1
and use the left-hand tool bar to make changes. These
changes will be reflected in the feedback display on
Monitor #2.
Below is a list of several
recommended video cards which provide the dual display mode feature on
desktop
PCs that have an AGP video card slot.
Check with your computer supplier or on-line suppliers such as
compuplus.com, compuvest.com, Spartan Technologies, etc.
VisionTek XTASY GeForce4
MX440 64MB DDR AGP
Model 30001520
Approx. price: $120.00
Matrox
Millennium G550 AGP 32MB DDR 2D/3D Dual DVI
Model G55+MDHA32DR
Approx. price: $120.00. NEEDS
ADAPTER.
Matrox
Millenium G450 Dual Heal 32MB
Model G45+MDHA32DLXB
Approx. price: $90.00. NEEDS
ADAPTER.
ATI Radeon
VE AGP 32MB Dual Display VGA & DVI
Model 100430119 Approx.
price: $60.00. NEEDS ADAPTER.
VIDEO CABLE ADAPTER: DVI 24-PIN MALE TO VGA 15-PIN FEMALE.
Return to Index
Choosing a Color Scheme
Physiolab has been designed
to mirror the colors of your chosen desktop color scheme.
If you wish to change the color scheme for
all of your Windows applications including Physiolab, right-click on an
icon-free area of your desktop and choose Properties. Click on the Appearance tab. Then click on the down arrow in the “Color
Scheme” box and select a color scheme to see a preview of it in the
Display
Properties window. When you find a color scheme you like, click “OK”
at
the bottom of the Properties window.
NOTE: While you can alter the Windows Color Scheme somwhat, you can not
alter signal colors or graph colors.
Return to Index
Overview of the Software Interface
This
section is an overview of the toolbars and signal controls common to
all
Physiolab applications.
Starting an Application
To
begin running an application click on the USE3 Physiolab icon
from
your desktop. The program will load to the
Main Menu.
Click on Select
Client. Select a client name in the
“Select Client” Window. Recorded session data will be saved under this
client
name. Click the “Select” button. If you
skip this step, you will be prompted later to select a client if you
record
data. (For information on adding new
clients, editing, and deleting client records, please see the “Adding
and
Deleting Client Information” section.)
Next, click on Run
Session. This will bring up the Select Application window.
Make sure that the J&J device
that you
have connected to your computer is selected in the “Hardware Type”
drop-down
box. Each hardware type has its own
unique application list. Click on the
name of the application that you wish to run.
Click Start.
If you receive an error
message, click OK, check the
connection to your computer, then retry. If you have trouble starting a
session, please refer to the “Troubleshooting” section in your Hardware
Guide.
IMPORTANT:
Once you have started an application you must exit it using the arrow
in the
lower left-hand corner, NOT the red
X in the upper right-hand corner.
Ensuring a Good Connection
In every application session, the first screen you will
see is the
Sensor Test Screen. Hook up the sensors using the
picture as
a guide.
Please refer also to your
Hardware Guide for information on cable/electrode configurations for
your
device and connection tips for different modalities.
The Sensor Test Screen
displays the impedance of each EMG/ECG/EEG electrode as a bar graph. The digital bar values are in K Ohms (K=1000
Ohms). J&J’s impedance testing
feature is important because it enables you to determine the quality of
your
signals. High impedance levels allow environmental electrical noise to
contaminate your signals. Values in the
green range indicate optimal functioning, yellow indicates marginal
functioning
and red indicates that the connection is inadequate and needs attention. (Please see the “Impedance Testing” section
of your Hardware Guide.) The bar colors are intended as guidelines only. The values needed to ensure good signals are
dependent upon the amount of electrical noise in your location and the
signal
level of interest. (For more
information please see the “Artifact Detection” section in your
Hardware
Guide.)
Signal bars for modalities other than EMG/ECG/EEG display
green for normal operation, yellow for needing adjustment, and red for
disconnected or broken.
For devices powered by
batteries, a digital battery indicator display near the bottom of this
screen
tests and displays the remaining voltage.
Return to Index
Checking For Good Signal Data
Click on Check Signals.
This screen displays all signals available in the application and
currently
being acquired so that you can inspect them to verify that they appear
in
normal ranges. It is not intended for
feedback. In Applications with EMG or
ECG, it includes a frequency spectrum (FFT) display which is
particularly
useful for detecting environmental electrical noise.
(Please see the “Artifact Detection” section in your Hardware
Guide.) A Check Signals 2 screen provides a screen without the FFT
display for
slower computers.
The Sensor Test and Check
Signals screens are unique to J&J and are designed to ensure the
quality of
your data. You should check these
screens each time you start an application.
Return to Index
Select A Protocol - Signal
Control Windows
Click on Select Protocol.
This screen displays many Signal
Control Windows, one for each signal eligible for feedback. Use
this Screen to set up which signals will be used for 'feedback and
which signals will be excluded. If you are setting up biofeedback for
Respiration, look for a Signal Contol Window like this:
Each Window has Activation controls on the left and Direction controls
on the right:
ACTIVATION BUTTON: Signal ON/OFF:Here is where you select or
de-select a signal for feedback.
DIRECTION
BUTTON:
* If you have de-selected this signal, the Signal ACTIVATION BUTTON is OFF, and
the Threshold Direction Light on the upper
right will turn Grey.
* If you
have selected this signal,
the Signal ACTIVATION BUTTON is ON,
and the Threshold Direction Light on the upper
right will turn Green or Red.
Adjust the DIRECTION BUTTON. Select 'On Above
Threshold' or 'Off BelowThreshold'.
The Threshold Activated light on the upper right will turn Green
if the Threshold condition is met, and Red if not achieved.

There are other buttons on the Signal Control Window. They will be
explained in the Signal Threshold section.
Return to Index
Select A Protocol - Pre-Set Protocols
Pre-set protocols are also available in the application so that you can
select the combination of signals to activate for feedback from a list.
Click on this icon on
the lower toolbar.
The Export Settings Windows opens.

Highlight a menu item and select Import.
NOTE:If this screen is blank, no protocols are available.
Return to
Index
Select A Task- Task
Control Buttons
This screen selects a task or tasks from
big buttons on the screen.
For example, a button may say, 'If
EMG is ON, Select EMG Task.'
Once a Task Button is selected, several events occur:
* The screen changes. This new screen is either a specific 'feedback
screen task, or a data entry screen.
* Depending on the Task, the program then pauses for you to enter data,
then records.
* Recording is set up by the Task. Tasks can vary from 30 seconds
to 30 minutes.
Timing is changed by toggling the Select Time Button on
most screens.
Return to Index
Toolbars Overview
Signals and Session Properties are controlled by Toolbars.
The top Screen
Select Toolbar provides buttons for choosing a specific display
screen.
The left edge Signal
Control
Toolbar provides buttons to change the display characteristics.
The Session
Control Toolbar at the
bottom of your screen provides controls for specifying tasks, recording
data,
marking events in the data, and for invoking various screen features
such as
screen freeze and sweep reset.
The Session
Status Bar, located below the Session Control Bar, indicates
elapsed and
remaining times for both the current task and the entire session.
The Task
Progress Bar It is a graphical
representation of the numbers in the Session Status Bar.

VIEW TOOLBARS
You can increase display
screen area and minimize distraction by hiding toolbars that you are
not
using. Simply click View, then
click next to the names of the toolbars you wish to hide. This will
remove the
checkmark next to the toolbar name and make the toolbar disappear from
view. You can retrieve toolbars
individually by again clicking View, then clicking the names of
the
toolbars you wish to reinstate OR you can retrieve all of the toolbars
at once
by clicking All Bars On.
Selecting Display Screens
The top Screen Select Toolbar lists
all of the
available display screens in an application.
Use this toolbar to select a specific display screen.
The screen button which is
highlighted shows the name of the current screen being displayed. Most
of the
top buttons have pull down menus (as shown above) of alternate screen
choices.
Clicking on the down arrow next to the screen name activates this pull
down
menu. Whenever you select a screen from the pull down menu, it is
assigned to
the top menu button. This ensures that
your favorite screen in each menu group will be only one click away. On slower computers, click on the drop-down
arrow and hold it down while mousing over the drop-down menu area, if
the
drop-down menu doesn’t at first appear.
Return to Index
Modifying Signals
-
Signal
Control Toolbar
Use the left edge Signal
Control Toolbar buttons to change display characteristics. These buttons will modify only one display or
signal at a time. To select an
individual display from a multi-display screen, click on the display.
The
display will become highlighted with light blue to indicate it has been
selected and that the Signal Control buttons have been assigned to it. Modify the signal using the icons to the
left.
If more than one signal
appears in the same display and the signal names are clickable, it
means that
separate scales are available for each signal and that signals can be
manipulated separately. Select the signal you wish to modify by
clicking on the
tiny colored square next to the name of the signal within the top of
the
display.
A white dot in the center of
the square indicates that particular signal is selected.
Use the Signal Control Toolbar to modify the
selected signal. If the signal names are not clickable, it means that
the
signals have the same scale and are locked together so that the signal
controls
modify all signals together.
The Signal Up button
and
the Signal Down button
are
used to move the signal line higher or
lower on the display.
The Auto-offset
button
toggles
on and off the Autocenter feature
which automatically keeps the signal centered on the screen. To manually adjust the signal higher or
lower on the screen, click Auto-offset OFF and use the above Signal Up
and
Signal Down buttons.
The Gain Up button
is
used to increase signal size and
sensitivity. Spaces between scale units
increase indicating greater signal sensitivity.
The Gain Down button
is
used to decrease signal size and
sensitivity.
The combination of Gain Up and Gain
Down plus the
Signal Up and Down arrows allow you to size and position the signal
anywhere on
the display.
The Autogain button
toggles
the Autogain feature on and off. For your
convenience you can leave Autogain
on or you have the option of turning this feature off and manually
adjusting
the gain according to your preference.
After you click Autogain on, wait up to 30 seconds for the
signal to
self-adjust.
Turning the Autogain or
Autocenter features on or off for one signal will not change the
settings for
the other signals on the screen. Each
signal’s settings are adjusted individually.
The Faster Graph
and
Slower Graph
buttons
are used to increase and decrease
sweep speed for all signals except FFT.
This increases/decreases the time span that can be displayed
within a
single window.
For FFT displays the Faster
Graph & Slower Graph buttons change the frequency span of the FFT
displays. It is recommended that you
leave FFT displays at the maximum frequency scale.
The Averaging Interval
button
changes signal appearance. This button
brings up the Select Averaging
Interval window which allows you to increase or decrease the averaging
interval
by seconds or fractions of seconds. Increasing the averaging interval
smoothes
the selected signal. Decreasing the
interval makes signal detail more visible.
DO NOT change the default settings for RAW signals.
The Zoom button
is
used to magnify any individual display to
fill up the entire screen. Simply click
on the desired display to select it, then click on the Zoom button. To return to the original display, re-click
the Zoom button. This feature is
particularly useful for simplifying feedback to one signal, then
toggling back
to a multi-signal display.
Clicking on the Adjust
Threshold button
is
enabled if “THR” is in the signal name of
one of the available signals for a particular display.
This button brings up the Threshold Control
window. An empty checkbox indicates
that Auto Threshold is OFF. To turn this feature ON, click in the
checkbox next
to “Enable Auto Threshold”. To manually set the threshold, click on the
check
mark next to “Enable Auto Threshold” in order to deselect it. Click on the Down/Up arrows next to the
“Level” box or type a numerical value in the box. Close
the control box by clicking on the small x.
Alternatively, you can drag
and drop the threshold line on the display screen.
Place the cursor in the display area and click the mouse once to
activate the drag-and-drop feature for that display. As you move the
cursor
over the threshold line, a hand appears. Hold down the left mouse
button when
the hand appears and drag the threshold line up and down on the screen,
releasing
the button at the point where you want the threshold line to stay.
The Show/Hide Signal
button
brings
up a control box to turn on or off the
display of a signal on the selected display.
Simply click on the check marks of each signal you wish to hide. To make the signals reappear, again click on
the Show/Hide signal icon and click on the empty boxes next to the
names of the
signal(s) to you wish to re-display. Close the control box by clicking
on the
small x.
The Select Sound
button
brings
up the Audio Control Window. Please see
the Audio Features section for
information on the use of audio controls.
The Sound button
toggles
the speakers on and off. This button must
be down in order for Media
Player feedback screens to work.
The Modify Settings button
is
used to specify numerical parameters, such
as breathing rate, duration of inhale/exhale and hold times for the
Breathing
Pacer or to specify the degree of smoothing (signal running average)
using an
EMG filter. It is also used to set time over threshold discreet reward
feedback
.
The Select
Second Screen
button
for Dual Screen Mode will open a
window with a list of available second screen displays.
To select a feedback screen to appear on a
second monitor, click on a display name and click APPLY.
You can run Dual Screen Mode IF your
computer is Dual Monitor enabled AND the USE3 software is set to Dual
Screen
Mode AND a second screen display is available in the application you
are using
(For more information see the section in this manual on setting up Dual
Screen
Mode).
The Show/Hide
Subwindows
button
allows you to hide and restore
subwindows in “All Signal” displays. This allows you to custom
configure a
display of any combination of the available signals.
With the Game
Files
button
you can select still and animated
images when you are on a game screen.
The Task
Averaging
button
opens a window that lists the mean
average for each signal and each task as it is recorded.
If any of the above buttons are
turned off they will
appear darker and less colorful. This
indicates that the button’s feature is enabled for that display, but is
merely
OFF, and the feature will be available if the button is clicked ON. Not all features are available for all
displays. Decisions have been made as
to which features are feasible and/or useful for each
display. If a feature is
not available for a particular display, its button will appear grayed
out
instead of “off” (darker).
Return
to Index
Session Control Toolbar
The Session Control
Toolbar at the bottom of your screen provides controls for
specifying
tasks, recording data, marking events in the data, and for invoking
various
screen features such as screen freeze and sweep reset.
Accessing Online Help
Click on the help icon
near
the bottom of your screen to access a
hyperlinked Table of Contents to an electronic version of this manual.
Exiting
a Session
Exit each session using the
arrow in the lower left hand corner
In order to review data, generate
reports, export data into a database, manage client or session data, or
to quit
the program, you must first exit the session.
In order to exit the software both the Pause and the Record
buttons must
be OFF. If you have recorded data, the Save Session Data window will
automatically open when you exit the software (See the Saving Session
Data
section).
Return to Index
Saving Signal Settings
Settings from the Signal Control Windows can be saved
[exported] an recalled [imported].

This is how you save and recall these settings.

The EXPORT
Settings
button
brings
up the Export Settings Window which
allows you to save all signal adjustments and settings as the defaults
for
future sessions.
If you wish to use the current settings, type in a protocol
name, and click EXPORT to save.
Many setting combinations may be saved for
each application. To recall a
previously saved protocol, click the Import
Settings button to reset the controls in the Signal Control Windows.
Return to Index
Reset and Freeze Features
The Reset Screen button
refreshes
all of the displays on the
screen. Resetting the screen has no
effect on data recording.
The
Freeze Screen button
freezes
all screen displays to allow for
discussion of the signals with the client while USE3 Physiolab
continues to
monitor and record data.
The AutoFreeze button
freezes
all graph displays when the selected
signal display reaches the end of its signal sweep.
This does not affect data recording. To
restart the signal display, click the Freeze Screen
button.
The
Pause Game Windows button
pauses
certain game and feedback screens
without affecting data recording. To
resume the game, re-click the button.
The Pause Task
button
pauses
both the data recording and the
feedback screen. The Unpause Task
button
resets the feedback screen and resumes recording of the current task.
The Pause Game
button
does
not effect recording. It merely freezes
the feedback screen while
the underlying data recording continues.
Return to Index
Printing Screens During the
Session
The Print Screen
button
allows
you to print the current screen.
Click this icon and wait a few moments for the screen image to appear.
Click
“Print” in the top menu, modify printer settings as needed, and click
OK. You are given the opportunity to abort
without printing by selecting “Close” instead of “Print”.
Recording Data
The Record button
toggles
recording on and off.
The Pause button
pauses
data recording. Re-click it to resume
recording.
The Menu button
allows
you to specify the erase time interval
invoked by the Record Backup button.
The Record Backup
button
erases
a specified time interval of recorded
data, and records over the erased portion so that the data record is
seamless. You can change the Backup
interval using the Menu button.
The Task
Averaging
button
opens a window that lists the mean
average for each signal and each task as it is recorded.
It verifies what is being recorded.
Return to Index
Event Marking
The Event Mark button
enables
you to add data markers into the
recorded data when significant events occur during the session. The button pops up a window into which you can
type a name or brief comment which is inserted with the marker into the
data
record.
NOTE: Event Marking is also done as an event on a window, such as a
Stress Scale.
Return to Index
Setting Up Tasks
Tasks are data recording
intervals which can be named to enable interpretation of session data. Task interval marks and color-coded time
interval bars are inserted into the graphic record.
Task labels are inserted into the recorded data file.
Specifying
Tasks Manually
To manually set up tasks,
click on the Select Task icon
in
the bottom Session Control Toolbar.
In the Programmed Task Schedule window,
you can
select different task labels manually at whatever time intervals you
wish. In
the lower left-hand corner click Manual Task to create a check
mark.
Click on the desired task name to highlight it, then click Run
Selected Task. The task name will
appear in the task progress bar at the bottom of your screen and will
be
inserted into the recorded data. When
you are ready to change tasks, repeat the above steps, highlighting a
new task
name. Click the record icon to start
recording the task. You can change the
task name while recording or you can opt to click the Pause button
between
tasks.
Running Preset Programmed Tasks
Click on the Select Task
button
in
the bottom Session Control Toolbar to
bring up the Programmed Task Schedule window. In
the Select Schedule drop-down box select a Task
Schedule name.
The
list of tasks in that schedule will appear in the large task status box. Click on a task name in the list and
click Run Selected Task.
Scheduled tasks are set to either progress
on to the
next task immediately or to pause before starting the next task. Check marks will appear in the task list
next to each task that has been run and recorded in the session. If a preset task stops recording and the
green “Pause” status message is flashing, you can opt out of the timed
pause by
clicking the Pause ON/OFF button to immediately run the next
task.
To manually pause while a task is running click the Pause
ON/OFF button. Re-click the Pause
button to the resume the
task.
To stop a task before it is finished and move on to the
next task
or to skip to a later task, click Record ON/OFF, then the
Select Task
button. Click on the name of the task you wish to run, then click Run
Selected Task.
To cancel a task schedule before it is finished and exit
the session,
click Record OFF. Then exit the
session using the lower left-hand exit arrow and either save or discard
the
data.
When a schedule of tasks
finishes running, the final task will remain paused until you click the
exit
arrow.
You
may run each task once per session. To
maintain data integrity, the data must then be saved and the session
exited
before you can run the same tasks again.
Customizing Programmed Tasks
You
can create multiple task schedules with unique customized tasks in each.
The
easiest way to start is to modify an existing task schedule for your
own unique
needs. When you finish modifying tasks
as outlined below, be sure and
click Update Schedule
to save all of your changes.
To begin, first click on the
Task
button,
then click Edit Mode On.
Click
the Select Schedule drop-down box and click on the task schedule name that most closely matches the list of
customized tasks you wish to create.
You
may move tasks up or down in the list, add new tasks to this list,
delete tasks
from the list, and modify the remaining tasks.
Each task must be customized individually. (If the task you want to
modify has already been run during the session the Task Properties
button will
be grayed out and you must exit the session and re-enter a new session
before
modifying the task.) Highlight the name
of the task you wish to customize, then click Task Properties.
Click Advanced
to expand the window.
In the Task Properties
window you can edit the name of a
task by simply typing over the existing task name (TIP: This task
name may
function as an instruction to the client since the task name is
displayed at
the bottom of the feedback screen) .
“Category” is an optional sort field not used in the current
USE3 View
function.
The Start Screen drop-down
box should be left at NONE unless you want the program to jump to a
pre-selected screen at the beginning of the task.
If you do wish to jump to a particular
screen, select
the screen name in the
Start Screen drop-down box.
Specify a Task Record Time in seconds.
Choose “This task bar only” to
display the progress bar for this
single task across the entire width of the screen.
Choose “All task bars” if you wish the progress bar to show
progress through all of the tasks in the task schedule.
You can loop
through a single task or through a series of
tasks a specified number of times. Check Loop then select
the name
of the task that you wish to have as the start of your loop. You may only loop back to tasks in the list
that are above the currently selected task. (NOTE: You will need to
click the
drop-down arrow and the tiny scrollbars to see all the tasks in the
list.)
Type in a Number
of Loops to indicate the number of times
you want the program to loop through the tasks before it moves beyond
the loop
to the next task in the list.
If Pause
Record
On Start is unchecked, recording will
automatically start at the beginning of the task and continue for the
number of
seconds you specify in the top right Task Record Time box.
You may set a
pause
at the beginning of the task and
let the task autorun so that the task begins recording after
the time
interval that you specify. Check Pause
Record On Start and specify a
number of seconds in the Pause and Text Time box in the
middle of
the window. If you wish to manually
start each task, you will want to specify a long pause time at the
beginning of each task. When you run
the task, click the Pause ON/OFF button to manually override
the pause
and begin the task.
Return to Index
Programmed Tasks:
Instructional Text Boxes
If you wish for a text
box with instructions to
appear at the beginning of the task, you can select one of the text
instruction
files in the Select Name drop-down box.
Click Test Text to position and resize the window and to
view the
text. USE3 Physiolab comes with a large
selection of instruction text files but, if you wish to add your own,
you can
create text files and save them into the C:\Program Files\J&J
Engineering\Physiolab\Text
folder. These files must be in .rtf
file format and must have unique file names.
If you wish to change the wording of J&J’s instruction
files, click Edit
Text after selecting the filename that you wish to edit. Be sure to click Save As and change
the file name so that your customized files will not be overwritten
during
software updates.
Check End
text
window on pause or record click if you
wish the test instruction window to close when recording begins
(either
an autorun pause times out or you click Pause ON/OFF to manually
override the
pause).
If End
text
window on pause or record click is
unchecked the text window will appear during the entire length of
the task. NOTE: You must click
Test Text and resize
the window in order for the text to display properly.
If you wish a
task
to be instructional only, with no data
recording, check the Record Off box.
Return to Index
Programmed Tasks: Audio Files
Physiolab comes with
many audio files which you may elect to play when the task
starts, when
recording starts or at the end of the task.
Text Window Start
files are
alert beeps or verbal instructions up to one minute in length. To play an audio file at the beginning of
the task, check Text Window Start and select an audio file name in the
drop-down box to the right.
Record Start audio files may be used as alert tones or
as
stressors (i.e. baby crying) in a stress profile task.
To play an audio file at the beginning of
recording, check Record Start and choose an audio file name in the drop
down
list.
If Pause Record on Start
is checked, you may
specify both
Text Window and Record Start sounds since the text window will appear
at the
beginning of the task and recording will start later.
If Pause Record on Start is unchecked, then the text
window will display at the same time that recording starts and only the
Text
Window Start audio file will be heard.
Task End should be used with care.
You may not have both an ending audio file for the current
task and a beginning audio file for the following task. If you specify both an ending sound for
the
current task and a beginning sound for the following task, then the
ending
sound file will be overriden by the following start task audio file.
The easiest way to become
familiar with USE3 Physiolab audio files is to go to the C:\Program
Files\J&J Engineering\Physiolab\Sounds folder and double-click on
each WAV
format audio file to play it. The subfolders are sounds that are used
in game
feedback screens and are not available as task sounds.
You may add your own files to the Sounds
folder. They must be WAV format audio
files and have unique names so that they will not be overwritten during
software upgrades.
Creating Your Own Programmed Tasks
Start by clicking on the Select
Task icon to open the Programmed Task Schedule window. Click Edit
Mode
On. Click in the top Select
Schedule box and type a new unique name for your task.
(Naming Tips: This name cannot be edited so
choose carefully. If you want a particular Schedule to appear at the
top of the
list keep in mind that schedule names are arranged alpha-numerically in
ascending order.) Click Add Task. Then follow the same process for each
task as outlined in the section on Customizing Programmed Tasks above. If you wish to add a new task in the middle
of the task list, click on an existing task name above which you want
the new
task to appear. Click Insert Task and proceed to create the task
in the
same manner as for adding a new task.
After you
have
specified settings for each task, click OK
in the middle of the Task Properties window.
When you have finished setting up all tasks, click Add As
New
Schedule in the Programmed Task Schedule window to save your
settings.
Saving Programmed Tasks
After you finish modifying
or creating each task you must click OK in the upper portion
of the Task
Properties box. If you click Cancel or
close the window with the X your settings will be lost.
If you are finished modifying tasks for an
existing task schedule, click Update Schedule.
If you are finished adding new tasks to a new Task Schedule,
click Add As New Schedule at the
bottom of the Programmed Task Schedule window.
The settings for Task
Schedules will be lost when you upgrade your software since they are
part of
the updated application. You can keep your current (older)
version of the application with your
customized tasks by going to the C:\Program Files\J&J
Engineering\Physiolab
folder and renaming the .mdb file for each application you wish
to save. These renamed customized
applications will
be not be updated. When you start USE3
Physiolab and click Run Session, two identical application names will
appear in
the application list. Find the one with
your customized tasks and run that one until you have time to customize
your
newer version of the software.
We recommend that you print
screen captures of Task Properties windows for all of your customized
tasks so
that you can re-create them in the upgraded software. (Instructions:
Open the
Task Properties window for each task and click Alt+Print Screen
on your
keyboard, then open Start> Programs> Accessories > Paint. In the Paint program click Edit,
then Paste. Click File,
then Print.)
Return to Index
Saving Session Data &
Verifying Client Names
If you have recorded any
data, the Save Session Data window will automatically open when you
exit the
software (using the lower left-hand Exit Arrow). You
have the option of discarding the data or saving the data to
a specific client name file. The names
displayed in the dropdown box are alias names.
You may select any alias name in the drop down box.
In order to verify that the alias refers to
the client that you intend, click Verify Name.
If you did not select a
client name before you ran the session, you can click New Client and
enter client information after exiting the session.
For directions on entering client data see the “Managing Client
Information” section)
Signal names that are
checked will be saved. You may uncheck
those
signals that you do not wish to save.
Click Save to Database. After saving or discarding data, click the X
to close the Save window.
If you click Export before closing
the Save window,
you have the option of exporting the current session data to Excel or
to
another database as described in the Exporting Data section.
Return to Index
Managing Data
If you are running a
session, stop data recording, exit the session to the Physiolab Main
Menu and
click Manage Data.
|
In the View
Session
window, select a client alias in the top drop-down box. If you wish to
see the client name, click Verify Name. Select
the application you used to record data in the second drop down box,
then click on a session date and time to select it.
At this point you can view data and generate reports, export data to
Excel or another database, delete data or archive data.
(Troubleshooting
Tip: If the session data you are
looking for is not present in the session window, it was probably
collected under a different application name. Select a different
application in the second drop-down window.) |
You can View, Export, ot Delete the Data.
Deleting Session Data
If you wish to discard
session data follow the steps above to get to the View Session window,
highlight the session date and time, then click Delete. To
delete an
entire client record including all session data for that client, see
the
“Adding and Deleting Client Information” section.
Return to Index
Viewing Session Data and
Generating Reports
The
View function
gives you the ability to review session
data in a graphical environment, replay signals, print screens and
generate
reports. Follow the instructions for selecting session data under
Managing Data
section above. Then click View.
If a blank window comes up, click on it to view.
Report Screen Names are shown on the top Toolbar.
Saved
Signal data is shown within each graph. Signal properties are
manipulated by HIGHLIGHTING the signal and using the left toolbar
incons.
The sweep time is changed using the horizontal arrows.
A new signal bar is shown on the bottom.
Return to Index
Changing Signal Views
In the Session
Review Window, you can replay the
session signals, selecting and adjusting the signals using the buttons
in the Signal
Control Bar in the same way as if you were running a session.
If
more than one
screen selection button is present at the
top of the window, you can use them to select the signal grouping that
you wish
to view.
If the signals are
against the top or bottom of the display
area, you must select the signal by clicking on the colored box in
front of the
signal name, then use the Signal Up
or Signal Down
button
to
manually adjust the height of the
signal. (You must first turn
Autogain
and Auto-offset Off.)
Using the Faster
Graph button
and
Slower Graph button
you
can change the amount of session time
displayed within the window, from focusing on a small segment or single
task to
condensing an entire session onto one screen.
Use
the Show/Hide
Signals button
to
zero in on just one signal or any
combination of signals.
With the
Averaging
buttons in the bottom Review
Session Bar, you can increase or decrease the smoothing of the
signal.
Click multiple times until you reach the desired degree of smoothing.
Use
the Rewind,
Scroll Forward, Scroll Back,
and End buttons

to
navigate through
the recorded session.
You can hide
and
re-display:
§
Task Markers with the Task
Mark button
Toggles the Pre-set Task Markers in
there are any.
Each application that has a Task Button also has a Task Marker.
It identifies time segments using a vertical line..
§
Event Markers with the Event
Mark button
Event Markers are being replaced by the
Event Marker button in experimental application.
§
Task Bar with the Task
Bar button
.
Toggles the Pre-set Task Markers
in there are any.
Each application that has a Task Button also has a Task Marker. Here
the button toggles the label, such as 'Stroop Test'.
§
Digital Cursor with the Digital
Cursor button
.
When the Digital
Cursor is toggled
on, you can click anywhere in the display to create a vertical cursor
line. The signal values at the point
where the cursor bisects each signal line are displayed next to the
signal
names as shown below:
Return to Index
Printing Screen Images
When you have adjusted the
signal displays as you want them to appear, click the Print Screen
button
in
the bottom toolbar. Wait a few moments for
the screen image to
appear. Click “Print” in the top menu, modify printer settings as
needed, and
click OK. You are given the opportunity
to abort without printing by selecting “Close” instead of “Print”.
Saving Screen Images
In addition to printing a
screen image, you have the option of saving a screen image as a
graphics
file. Click on the Save Picture button
.
In the window that appears, browse to the
location where you want to save the image file in the top drop-down box. Give the file a unique filename and click
OK.
Printing Quick Reports
You can fine tune your data
views by adjusting the signal displays then printing Quick Reports. First select a client, an application, and a
session as described above in the ”Managing Data” section.
Manipulate the signals to obtain the screen
view to be inserted into your report as outlined above in the “Changing
Signal
Views” section. Then, to print a
pre-formatted Quick Report, click on the Generate Report button
button.
This brings up the Generate Report
window. Accept the defaults or use the checkboxes to designate the
types of
information to be included. Select a
different averaging interval or accept the default.
Accept the Average Only default or choose Basic
Statistics. Average Only includes
mean averages for each signal over the selected interval.
Basic Statistics includes standard
deviation, minimum and maximum for each interval.
Click “Print Quick Report”.
Return to Index
Creating Excel Reports
within View
Tis is a special button for Excel users.
Click on the Generate
Report button
.
Accept the default or select the desired
averaging interval for your report, click
Excel Report.
You will see a pre-formatted
report generated in Excel with the current Session Review Screen image
inserted
into the report. You will be prompted to save your report.
Browse to the your desired folder location
and give your Excel report a name, then click “Save”.
To view or print your report, insert text, or chart data you
will
need to re-open the report from the location where you saved it.
You can insert rows and
merge cells to create text boxes for expanding and
personalizing your
Excel report. Click on the row number to the left above which you wish
to
insert rows of text. Click on Insert>Rows. Repeat this to insert as
many
rows as you wish. Left-click on the cell where you want the top left
corner of
your textbox to be. Holding the left mouse button down, drag the cursor
down and
to the right to highlight the number of rows and cells desired. In the
top menu
choose Format>Cells. Click the Alignment tab. In the “Horizontal”
drop-down
box choose “Left”. In the “Vertical”
drop-down box choose “Top”. Check “Wrap Text” and “Merge Cells”. Click
OK. You
now have a text box in which to insert paragraphs of text.
It is best to start with a text box slightly
smaller than you think you will need.
If you need a larger box, click in the existing text box and,
holding
the left mouse button down, drag the cursor to highlight cells to the
bottom
and/or right. . In the top menu choose Format>Cells. On the
Alignment tab,
click “Wrap Text” and “Merge Cells”. Click OK.
You can use the Excel Chart
Wizard to create graphs. Select single columns of signal data
because
Excel can only chart one scale at a time.
Click the
Chart Wizard Icon in the top menu.
Select the
type of
chart you want. You can try out different
options and view them in the preview window to see which displays of
the data
are most effective. You may find that your sample rate is too high or
too low
to create a meaningful chart. You can
re-send the report to Excel at a different sample rate using the Excel
Report
button, resave it to a new filename, and re-chart it using the Excel
Chart
Wizard.
You can save your Excel
report using the File>Save menu commands
and/or print out your report using the using the File>Print
commands
in the Excel menu bar.
Return to Index
Exporting Data to Excel or Your Own
Database
Exporting data does not
affect the original data record. It
only exports copies of the data. To
export data, first follow the instructions for selecting session data
under the
Managing Data section above. Then click
Export.
In the right side of the
Export Dialog window, uncheck any information you do not wish to
include.
In
the Update drop-down box, choose the update rate, which is the
averaging
interval time in seconds (decreasing the rate reduces file size and
makes
charting the data more feasible).
Select either Average Only or Basic Statistic (described
in the Quick Reports section).
If you don’t want to export all of the
signals, highlight individual signals and move them into the Saved
Signals box using
the left-arrow.
If you wish to export
to Excel, select Excel in the Export Type box
and-click OK. Click Yes in
the save prompt pop-up
window, browse to a folder location, type in a unique filename and
click Save. If you wish to include
screen captures in
your Excel report, or add textboxes or charts, you may want follow the
directions in the Creating Excel Reports section instead of exporting
data.
If you wish to have
Physiolab automatically generate filenames and save to a pre-selected
target
location, check the Autosave checkbox, then follow the
instructions below
in the Automatically Generating Filenames section.
When Autosave is checked
Excel will pop open and populate with data, then the save window will
briefly
appear and disappear as the file is automatically saved to your target
location, and Excel will close. It may
take a few minutes for these steps to execute, especially with large
files.
If you wish to export
to another database, select Asci in the Export Type box.
You must click Change
File to specify an initial target location. In
the pop-up window, browse to the folder into which you want to
save the data. You must replace the asterisk(*) with a unique filename,
then
click Open.
It is important to specify
which symbols your database will be looking for in order to interpret
divisions
between data elements. Click the radio
button next to the appropriate symbol combination in the “Data
Delimiter”
box.
Do not close the Export
Dialog window using the X in the upper right corner. The file will not
be saved
until you click OK in the Export Dialog window.
If you are using your own
database program, use the instructions supplied with that program for
importing
and manipulating data.
Return to Index
Automatically Generating Filenames for
Export
Data
Once you have used Change
File to specify an initial target location and filename, you can let
Physiolab
automatically generate a new filename and export to your initially
specified
location by default each time you export data.
When you click Export the next time, the target
location and
filename from your last export will be used again unless you
specify
something different. It is important to select Auto Increment
in the
Export Dialog window or the exported data will overwrite the previously
saved
file with the same filename. When Auto
Increment is checked, the number in the Auto Increment textbox will be
added to
your initial filename. When you click OK at the bottom of the
Export
Dialog window, the data will be exported and
the Auto Increment number will increase by one integer. You can start Auto Increment over at zero or
at any other number you choose by typing that number in the textbox to
the
right of the Auto Increment check box.
Even with Auto Increment
selected, you always have the option of manually specifying a
unique
filename by clicking on the “Change File” button, typing your new
filename,
then clicking Open. The number in the Auto Increment textbox will be
added to
your manually typed filename when you click OK in the Export Dialog
window to
export your data. This new filename will
become the new default template filename unless you change it back.
To automatically generate data export
filenames using
a client name as the filename template, both the Use client
name as
Template and the Auto Increment checkboxes must be checked.
Even if
you have exported data before using “Change File” to specify a
location, you
must specify an initial target location again when switching to using
the
client name as a filename template. When “Use client name as a
template” is
checked, the “Change File” button opens a window to select a file
export
location only, without a textbox for specifying a filename. Browse to the data destination folder of
your choice, then click OK to confirm your target location and
close the
window. Click OK
in the Export Dialog window to export your data.
For future data exports, once you have
selected a data session and clicked Export, your target location and
filename
will be automatically set and you will need to simply click OK to
export.
Return to Index
Archiving Session Data
When you
have stored ½ gigabyte of data, you will start getting
messages suggesting that you archive files.
It is important to respond to these prompts and archive data
since you
will not be able to access session data if the 1 gigabyte
maximum
database size
is reached.
All patient
data is
saved into the same Physiolab data.mdb
file. This data is coded, and requires a special
program to archive. That program is found within PhysioLab.
To archive
data,
click Manage Data on the main entry menu,
then click Archive on the far right.
The Archiving Widow is activated.
In the
Archiving
window, select a client alias name in the
Main Database drop-down box. Select an application that has data for
the client
in the top drop-down box. To
select a target folder location for
your archived files, click the small button to the right of the Archive
textbox. Either select an archive file
name in the Open window (if you have previously archived data) or
browse to the
folder location where you wish to create a new archive, type a new name
in the
“File name” box, and click Open. You
can choose to have a separate database file for each
client or you can archive files by time period (quarterly, half-yearly,
etc.)
Back in the
Archiving window, click on the date of a session
in the list to the left, then click the right arrow to move it into the
list of
files to be archived. You can
select
and move all files in the list or
you can pick and choose sessions by holding down the Cntl key
while clicking on those you wish to
archive. Finally, click Compact Databases to free up space in
the
Physiolab database, confirm by clicking Yes, and click Close
to
exit the window.
If you wish
to copy
a file into a new data location, while
leaving the original file in place, follow the procedure for archiving
files,
but click the Copy dot before you click
the right arrow to move files into the right side of the Archiving
window (This
does not free up space in the data.mdb file). Click Close to exit the
window.
At any time
you can
move archived files back into the
original Physiolab database to view
them. Follow the same procedure for
archiving files, but select archived sessions in the right side of the
Archiving window and use the left arrow to move them back into
Physiolab.
Return to Index
Adding and Deleting Client
Information

To add a new client, on the
main
menu click Select Client, then click New.

Enter the new client’s information. In the
top textbox you have a choice of letting the program generate an alias
name by
leaving the box blank or of entering your own alias name. Automatically
generated alias names consist of the first two letters of the last name
followed by the first two letters of the first name.
If two clients have similar names so that identical aliases are
generated, the program will add a number after the duplicate alias to
differentiate them. The alias name will
be generated when you exit the window by clicking OK.
To edit a client
record, click Select Client as above, then click Edit. Modify the client information as desired,
then click OK.
To delete a client including all session data for that client:
On the main Physiolab entry window click Select Client,
highlight the
client alias name, then click “Delete”.
Click “Yes” at the warning prompt to delete the record.
It is a good idea to click Edit
to verify that you have the correct client name before deleting
client
records.
For
information on viewing client data see the Managing Data
section.
Return to Index
Overview
of Screens and Displays
At the top of the session
window in each application is a Select Screen Bar with drop
down lists
of screen choices and hot key buttons that can change to
provide a quick
link to the most recently selected screen in each list.
Each Screen may contain a
single display which fills the entire window or multiple displays tiled
within
the window. A good example of a tiled
multi-display
screen is the Check Signals Screen in every application. Below is an
example
from the C2 EMG Application:
An individual display within
the window can be selected by clicking on it and the Zoom button can be
used to
fill the entire window with the selected display.
Each display may contain a
single signal or multiple signals. By
the light blue highlight color surrounding it, you can see that the
FFT_EMG A N
display has been selected. Since there
is only one signal reflected in this display, selecting the
display also
selects the signal, which can then be modified using the Signal Control
Bar. (See “Modifying Signals”). The
signals in FFT_EMG A N and FFT_EMG B N look quite different because
they have
been individually selected and their scales modified independently with
gain
and sweep speed controls.
The SR-C and TEMP-C displays
above each contain multiple signals sharing the same space in a
single
line graph. Since these signals are of the same type and share
the same
scale, the Signal Control Bar buttons will modify all of the
signals in
these types of displays together. For
example, if the SR-C display were selected and the Gain Up button were
clicked,
gain would increase for both signals in the display. Multiple signals
can also
share space in scrolling charts.
In other multiple signal
displays, signals of different types can share the same space
in a
single line graph or chart. Since only
one scale for one signal at a time may appear in this type of display,
you can
click on a signal name to change the display scale so that it matches
the
selected signal. Notice in the example
below that the HR signal has been selected, and that the scale to the
left is
in beats per minute to match the selected signal. Only the selected
signal will
be modified by the Signal Control Bar buttons. For example, to select
and
modify the Resp1 signal below, you would click on the small blue box
next to
“Resp 1”. Note that the second respiration channel is not being used,
but the
Resp 2 signal name is listed to show it is available. If a signal is
not being
used, as in the example below, or if it is accidentally adjusted off of
the
screen with the Signal Up/Down buttons, the signal is still present
scrolling along the edge of the display so that you can locate it. In the example below, the Resp 2 signal is
present as a tiny red square just under the Resp 2 name.
In contrast to this display,
in displays with the same kinds of signals sharing the same vertical
and
horizontal scales (as in the SR-C and TEMP-C examples above),
individual signal
names are not “clickable” and signals cannot be modified individually.
In still another type of
multiple signal display, each signal has its own line graph, with the
graphs stacked
parallel to each other and a common time scale along the
bottom. The EMG
A N and EMG B N signal graphs in the C2
EMG Application Check Signals Screen above are considered to be a
single stacked
graph display. Clicking on either
graph will highlight both, and the Signal Control Bar buttons will
modify both
signals together.
In
contrast to this notice the EMG A-N and ECG B (Raw) displays
below which are completely separate displays with independent
vertical and
horizontal scales.
The Resp1 + Resp 2 display
below is a slightly different variation of a stacked graph display in
which the
time scale is shared so that the sweep speed buttons will affect both
signals,
but the vertical scales are independent so individual signals
may be
selected and Signal Up/Down and Gain Up/Down controls are applied
separately
for each signal.
In the above overview of
screens, displays and the signal configurations they can contain, you
have seen
line graph displays, FFT displays and, in the screen immediately above,
a bar
graph display. The next section
provides an overview of these display types plus the other display
types you
will encounter in Physiolab.
For detailed descriptions of
the Sensor Test and Check Signals screens, please see
the “Ensuring a Good Connection” and
“Checking
for Good Signal Data” in the “Overview of the Software Interface”
section.
Return to Index
Display Types
The following display types
may fill an entire screen or may be tiled together in any combination,
depending upon screen space and the usefulness of correlating specific
displays
together. The same display type may have different signal, background
and fill
colors within different screens and/or in different applications.
Zooming In
Any of the displays
described below may be selected and the Zoom button used to fill the
entire
screen with the selected display for focused feedback.
Line Graphs
Line graphs emulate an
oscilloscope-type display, with one or more signals scrolling left to
right.
Since full descriptions were made of several types of line graphs
above, a
brief summary of line graph types follows:
- Single
line graph with multiple signals sharing the same space
- If
all signals are the same modality with vertical and horizontal scales
common to all signals, signal names are not clickable and Signal
Control Bar buttons adjust all signals together.
- If
signals are for different modalities with different scales, signal
names are clickable and signals are adjusted individually.
- Stacked
parallel line graphs
sharing a common time scale, with a single signal in each
- If
signals are for the same modality so that the vertical scales are the
same and the training purpose is to cause the signals to move in the
same direction, signal names are not clickable and Signal Control Bar
buttons adjust all signals together.
- If
signals are for the same modality but the training purpose is to cause
signals to move in opposite directions or the signals may be monitoring
two different individuals, signal names are clickable and the signals
are adjusted separately
- If
signals are for different modalities with different scales, signal
names are clickable and signals are adjusted individually
Pattern Displays
A pattern display is a
special type of line graph that contains a Pattern Signal. The Pattern
Signal
does not reflect physiology but, instead, provides a guide with fully
adjustable parameters set by the clinician.
Sharing the space in the graph with the Pattern Signal is a
signal
reflecting physiology. The goal is for the client/patient to modify
his/her
physiology to match the pattern signal as closely as possible. An example of a pattern display is the Breathing
Pattern display more fully described later in the “Respiration
Displays”
section.
Ghost Displays
The Ghost Display is another special type
of line graph,
which contains a single signal. With
each signal sweep across the screen, the previous sweep fades but
remains
visible for comparison with the new brighter signal sweep. You can see
in the
example below that multiple sweeps remain visible for comparison across
time.
Scrolling Chart Displays
Useful for showing trends over time,
Chart Displays
have a diamond-shaped tracer to the right which “draws” the signal as
it
scrolls to the left emulating a traditional paper chart.
Though the signals scroll in the opposite
direction as line graphs, the oldest signals appear to the left, with
the most
recent signals to the right, consistent with line graph displays. As with line displays, a chart display can
contain a single chart with a single signal, multiple signals within a
single
chart, with or without fill colors, or stacked signal charts as in a History
Chart example shown below. All
signals share a common time scale so that Faster Graph/Slower Graph
buttons
adjust all signals simultaneously. Vertical scales are individually
adjustable.
Fill Graphs
Fill colors are used in both
line graphs and in scrolling charts as shown above.
The fill colors used above are used to give the signal emphasis,
especially in reference to zero or some lower point in the scale. Fill colors can also be used as color
rewards for achieving threshold goals as in the Threshold Up, Threshold
Down,
and Threshold Up/Down Displays below.
The Threshold Up display provides a color reward when a
threshold is
exceeded. The Threshold Down display is
useful when the goal is to modify physiology by bringing the signal
down. Fill and background colors may vary
in
different screens and applications.
Return to Index
Threshold Up
Threshold Down
Threshold Up and Down
A dual-purpose display, which
provides
different fill colors above and below the threshold line.
It is structured like the Zero Point Fill
Display pictured below except that the reference point for the change
colors is
the threshold line instead of zero. It
can be used as either a threshold up or threshold down display, with the goal area color providing a reward
and the out-of-goal area color providing a reminder.
Adjusting Thresholds
The threshold signal name is
listed in the display but you cannot click on this signal name to
select it
because none of the Signal Control Bar buttons apply to the threshold
signal
except the Adjust Threshold button. In displays with thresholds,
AutoThreshold
is defaulted to ON so that it is at a known visible point in the
display. If you wish to disable
AutoThreshold and
manually set your own threshold, click on the Adjust Threshold
button
to
bring up the Threshold Control window.
Next click on the check mark next to “Enable Auto Threshold” in
order to
turn the automatic threshold adjustment feature OFF.
Click on the Down/Up arrows next to the “Level” box or type a
numerical scale value in the box. The
threshold line will move to point on the scale that you specify. Close the control box by clicking on the
small x. If you want to turn
the Auto Threshold feature back on, click on the Adjust Threshold
button, click
in the empty checkbox next to “Enable Auto Threshold”, then click on
the x
to close the pop-up window.
Another way to adjust the
threshold is to drag and drop the threshold line on the display
screen. First put the cursor in the
display area and click the mouse once to activate the drag-and-drop
feature.
Now as you move the cursor over the threshold line the arrow will turn
into a
hand. When the hand appears, click-and-
hold the left mouse button while dragging the threshold line up or down
on the
screen. Release the button at the point
where you want the threshold line to stay.
Return to Index
The following are fill graphs
without thresholds:
Differential Fill
The purpose of
the differential fill graph is to vividly distinguish between
the
activity of different signals.
Especially effective for differential muscle training.
Bottom Fill
The bottom fill graph is the same as above
but the color
of the signal with the higher value extends to the bottom line of the
graph.
Zero Point Fill
This type of differential
fill can be used in line graphs and in scrolling chart displays as
shown
below. The zero reference point is
useful for showing direction of change, for example in Temperature
warming/cooling and Skin Conductance response displays.
Rainbow Fill
Especially effective in
Temperature and Skin Conductance displays, Rainbow Fill graphs provide
incremental color rewards.
Specialty Fill Graphs
The following are special
purpose fill graphs developed for particular modalities. These displays
are
discussed in the sections below devoted to displays, settings and
parameters
specific to each modality.
Return to Index
EMG Displays:
FFT and 3D FFT displays are
discussed in the “EMG
Displays” section below.
HR and HRV Displays:
ECG FFT, DFT, VLF,LF,HF and SDNN displays
are described in the “HR and
HRV Displays” section below.
Bar Graphs
Standard Bar Graph
A
vertical
bar graph displaying live signal values, pictured below left.
Threshold Bar Graph
A
standard bar graph with an adjustable threshold line, pictured
below center. Threshold adjustment
instructions are the same as for line graphs with thresholds on page 21.
Zero Center Bar Display
Pictured to the right below,
the center of the scale in this display is zero, with negative values
below and
positive values above. The bar fills
either up or down depending upon signal polarity. It
is used for displaying signals, such as rate of change, that
can move in either positive or negative directions.
History Bar Display
30-sec averages are
displayed in this scrolling bar graph display pictured below in the
“History or
Trend Displays” section.
Numerical Displays
These digital displays
enable background monitoring of signals that are not the primary focus.
Return to Index
History or Trend Displays
History Bars
This scrolling bar graph
screen, with its combination of live signals and 30 second histograms,
is
useful for its ability to show trends over time. As
each signal completes one 30 second sweep, a bar is added to
the graph showing the thirty-second average value of the signal. We suggest that you leave the signal sweep
interval at thirty seconds in order to match the bar graph update. The
time
interval is the same for each bar graph display so that the
Faster/Slower Graph
buttons change all graphs simultaneously. You can select and eliminate
the
signal sweep line graphs in order to provide more screen space for the
history
bars. You can change the number of bars
displayed using the Faster/Slower Graph buttons and can even condense
the bars
to view the entire session on a single screen. This screen resets when
you
begin recording a session.
History Graphs
Stacked Line graphs containing signal
combinations
moving at a sweep speed too slow for feedback, but very useful for
seeing a
long time interval of the session and viewing trends over time.
History Charts
An example of the History Chart
Display is pictured
in the “Scrolling Chart Displays” section in the discussion of display
types
above. History Chart Displays are not
meant for feedback, however they are much more useful than feedback
screens for
evaluating overall trends. By repeatedly clicking the Slower Graph
button, you
can display ever larger time segments, even the entire session, in a
single
display window. Autogain is defaulted
to ON for this screen. You can manually adjust individual signal
magnification
by clicking Autogain OFF and using the Gain Up/Down buttons. This
screen resets
when you begin recording a session.
Return to Index
FEEDBACK
INFORMATION GRAPHS
Data from the Signal Control Window is displayed several ways.

This is a Signal Control Bar for EEG:

Success at achieving the biofeedback response is indicated by a scroll
bar.

When all criteria for a successful biofeedback response is met, the
START bar turns green.
Whn the START Bar turns green, the Score or On Target Bar advances one
step.
When the On Target Bar is full, it resetes an he LAPS Bar advances one
step.
FEEDBACK SUCCESS GRAPHS

The success data is saved in the session file for playback.
When the graph turns green, the feedback criteria were achieved.
Return to Index
EEG Displays
EEG Impedance Screen
This
first screen is
the hookup and electrode impedance
screen. The inputs to the EEG channels
are shown as A+ B+ C+ and D+ for a four-channel application. A four-channel application will have a
Reference 1 and Reference 2 which will be connected as the monopolar
reference. A two-channel application
will just have A+ and B+ and the references will be called A- and B-.
In a two-channel
application you have the option of switching between monopolar and
bipolar
configurations. This impedance test
screen is showing the live impedance of each electrode to ground. In additional to the digital reading in
K-Ohms there are red-yellow-green colored dots indicating: green - under 5K, yellow - between 5 and 10K, and
red -
over 10K. Instructional
information, if provided, appears in a panel to the right with Previous
and
Next buttons to cycle through multiple pages.
Some
applications
allow you to assign label names in this
impedance screen. In other applications
you will need to click Check & Label in order to assign label names. If the application allows, you can label the
channels with the site name as you are hooking up electrodes by
clicking the
site on the headmap, then clicking on the white space next to the
corresponding
channel to the right. After the
impedances are satisfactory, click Check & Label to check raw
signals and
change site names if required.
EEG Setup
When you are
satisfied with your label names and EEG
waveforms, click EEG Setup. This screen
is the primary control screen for all EEG feedback.
The top display is the raw EEG waveform for one, two or four EEG
channels. Below are the EEG band and
threshold control displays which will have between four and twelve
functionally
identical displays that can achieve very different effects depending
upon how
you set the adjustable features. Most
EEG protocols can be achieved by adjusting these individual bands and
thresholds.
Each EEG
channel
will have from two to four bands connected
to it. The bands are color-coded to
match the trace color of the channel they are connected to. They are further identified by name. For example, Band
2-A indicates that this trace represents the second band
connected to the A channel. This name
will be used in feedback displays and in recorded data.
The Band
ON/OFF button in the upper left corner of the band display controls
whether the
band is used as feedback. If this button
is toggled OFF, this band is not used for feedback and the top right
color bar
is grey.
In the three
sets of
Arrow Up / Arrow Down buttons to the left, the leftmost arrows
are used
to specify the frequency of the bottom edge of the band (HP), the
right-hand
arrows are used to specify the frequency of the top edge of the band
(LP), and
the middle arrows move the entire band up or down by equal increments
of .2
Hz. The number in between “HP” and “LP”
indicates your current setting for the band width.
The normal
convention for setting up the bands is to use
Band 1 for Reinforcement , Band 2 for Inhibit, and band 3 for artifact
inhibit. You should set up a minimum of
two Inhibit bands to eliminate artifact as follows:
1.
Inhibit settings
of 1-4
Hz for eyeblink and lead motion
attached to a frontally located lead and
2.
Inhibit settings
of 30-60
Hz for EMG attached to any site.
Return to Index
EEG Band Threshold Overview
The two
horizontal
lines in the graph area are the physical
positions of the EEG Band threshold. The
threshold lines do not move. Only the
scale of the display changes. The
controls on the right side of the band are for setting threshold. When the signal line crosses either of the
threshold lines the threshold state changes.
The effect of this change is determined by the position of the THRESHOLD DIRECTION button.
A green color
bar at
the top of the controls to the right of
the signal band indicates that your threshold target has been met. Any red bar on the screen will inhibit
feedback. When all bands are within
target for the length of time (step) that you specify, feedback is
given. Feedback will only occur if all
bars are
green or grey.
If the signal
has
been inside the threshold lines for some
time and it crosses outside either threshold line, the indicator will
change
state and hold state for .1 second regardless of what the signal does. This prevents the indicator and feedback
display from flickering. There is no delay when crossing the threshold
in the
success direction.
The number to
the
left of “SIG” indicates the amplitude of
the signal in peak microvolts. The number to the left of “THR”
indicates the
amplitude of the target threshold.
Below this is the percentage of the time that threshold is
achieved.
Threshold Above and Threshold Below Settings
If
the THRESHOLD DIRECTION
button is toggled on to “Above Threshold”,
the signal line must extend beyond both the top and bottom
threshold
lines in order for the green indicator bar to show that the target
threshold
for this band has been met .
If the THRESHOLD DIRECTION button is toggled
on to
“Below Threshold”, the signal line must
remain
within both threshold lines in order for the green indicator bar to
show that
the target threshold for this band has been met .
MAN vs. AUTO Threshold Mode
If “MAN” is
selected, the up/down arrows to the right of the
signal can be used to manually increase or decrease the threshold value. You can manipulate the threshold according
to the client’s signal amplitude to achieve a desired success average.
If AUTO is
selected
then the up/down arrows to the right
adjust the desired success rate percentage and USE3 Physiolab adjusts
the
thresholds to enable the client to achieve this target success rate. If Inhibit is selected, the success
percentage needs to be set much higher
than if Reinforce is selected in order to compensate for the dropout
delay.
Be sure that
the Score/On Target Bar at the bottom of the
screen is moving. If it isn’t moving,
you need to readjust the threshold controls appropriately for the
client so
that a reasonable success percentage is being achieved.
Threshold Average and Peak Average
At the bottom
of the
EEG setup screen two control functions
called Threshold Average and Peak Average.
The up-down
buttons
to the left of the “Thr Avg” changes the
AUTO Threshold update
rate. Altering the Threshold Average
changes how frequently in seconds the AUTO threshold is re-calculated.
120 seconds or more are used.
The up-down
buttons
to the left of the “Peak Avg” label
increase or decrease the
averaging function (smoothing ) of all signals.
Types of EEG Feedback
The EEG
applications
are set up to do two different types of
feedback with different audio feedback and display options for each
type.
The first is
traditional threshold-driven continuous
reinforcement with feedback consisting of MIDI synthesizer music,
recorded
voice or music WAV files, Media Player video files or music CDs. This type of feedback is controlled using
the Sound Select window described in the Audio Features section. To control feedback you can use individual
bands or you can combine all bands that are turned on into one signal
(called
“yoked thresholds”) by selecting “Audio All Thr Tune” or “All Bands”. This type of continuously running feedback is
turned on when all thresholds are grey or green and feedback turned off
when
any threshold turns red.
The second
type of
feedback is response-reward-timeout
feedback with a beginning, discreet steps, and an end, which is a
reward
period after a fixed number of steps.
Progress is measured in rewards per minute.
Rewards consist of pictures revealed in steps or specialized AVI
files that show one image frame per step.
Research
by Dr. Barry Sterman indicates that learning is superior with response-reward-timeout
feedback. It is now used exclusively in these applications.
Return to Index
EEG Feedback Score Bar
The
Score or On Target function
is designed to provide a discreet reward
(step) for holding a set of threshold conditions for an interval of
time. Examples
are: displaying a new portion of a larger picture puzzle
or
displaying
the next picture in a sequence of pictures.
The bar to the right of the word Score at the bottom of
your
screen
displays progress through completed steps
The number of steps represented in this bar is
determined by the
setting
you choose in the Steps to Score drop-down box in the Score Properties
window.
The Score
function
combines the threshold outputs from all
EEG bands into a single output. When
all band indicators are either green or grey (off), indicating a
threshold
success state, the Score timer is started.
Feedback is provided after a time interval that you specify. Each time feedback is provided the yellow
Score bar moves one step to the right.
The Score
function
is not used for continuous running type
rewards, such as cartoons. These
rewards are controlled by the Select Sound window described in the
Audio
Features section.
Return to Index
Overview of Picture-Reveal Games
Each
picture-reveal
game screen consists of a grid of
squares that can be uncovered to reveal a larger picture.
The picture is revealed in discreet steps
with the final step completing the remainder of the picture. The name of each game screen indicates the
number of squares in the columns and rows of the screen grid. Multiply the column number by the row number
to determine the total number of squares in the grid.
Knowing the
number
of grid squares will help you to set an
appropriate number of steps for the client to complete in order to fill
one set
(or image) for that screen. For
example, for picture-reveal type games the step number needs to be less
than or
equal to the number of grid squares. A higher step number reveals a
smaller
portion of each picture at each step. A
lower step number setting yields a higher reward rate because all the
remaining
squares are revealed on the final step.
In screens
with
fewer grid squares, more of each picture is
revealed at each step.
Screens
with a lower
number of grid squares are appropriate
for use with children, for example the 6X1 screen can be used for
simple race
games. For adults, you may want to use
art or travel photos with an 8X8 screen.
Feedback Time Settings for Games
To
modify feedback
parameters from the default preset
values, click Score at the bottom of the screen, then the Modify
Settings
button
in
the left toolbar.

In
the Properties
window you can set the number of
steps to complete a screen, specify time intervals, and setup feedback
sounds.
The top time
variable, Steps to Score, is the number
of success state intervals required to complete a set (fill the Score
indicator
bar). An appropriate number Steps to
Score is determined by the type of game being used.
Some games require a fixed number of steps to operate properly
and, in many cases, the number of steps will be preset by the game. We recommend you accept these default
values.
The Response
Time Per Step is the length of time in
seconds that all thresholds must be continuously green to generate a
step in
the game and move one step- increment on the Score indicator bar. The client must keep signals within
threshold parameters for the interval you set in order for feedback to
occur. Response Time Per Step is
normally set from .25 to .5 of a second.
The Timeout
Between Steps is the dead time after a
step has been completed.
The Timeout
between
steps is normally set from 0 to 1
second.
Timeout
After
Score is
the timeout after a set has
been completed.
Feedback Sound Settings for Games
The sounds
associated with Score are selectable from
pull-down lists and can be individually turned on and off for the
following
functions:
Start New
Game sound occurs once
only
at the
start of a Score set. An example would
be a game whistle.
Response
Start sound is played
any time
all of
the thresholds are simultaneously green with no time delay. It indicates that the threshold conditions
have been met. This sound should be
short (less than the Response Time Per Step).
Response
Complete sound occurs when
all of
the
thresholds have been held for the required time (Response Time Per
Step). This occurs at the same time that
the
feedback display is triggered.
Score sound occurs at
the end
of a Score set (Score Bar
filled - feedback display completed)
Return to Index
Selecting a Game Screen
The same game
screens are
available for single monitor display and dual screen display setups. If you are using a single monitor, select a
game screen from the drop-down menu in the top Screen Select Bar. If you are using a two monitors
click the Dual Monitor button
to
bring up the windows for selecting game
screens for the second monitor, highlight your screen choice, then
click Apply.
The Secondary
Screen
List contains all screens available for
display on a second monitor, including game screens, bar and line graph
displays, Media Player and AVI screens.
All of these same screens are available with single monitor
display but
the non-game screens are selected from other menus in the software.
Picture-Reveal Screen Types
The following
screens randomly reveal one square of an image
per step:
8X8 1
is
a
64-square grid with the Score bar along the bottom.
8X8 2
is
a
64-square grid with band output displays along the bottom showing band
1 from
channels A&B as reinforcement and band 2 from channels A&B as
inhibits.
5x4 is a 20-square
grid with
the Score bar along the
bottom.
13x4 is a 64-square
grid with
the Score bar along the
bottom.
The following
display screens reveal one entire picture or
one square of a larger picture per step in a sequential order. Vendors have developed specific games, such
as races, to go with these screens.
Contact J&J for a game vendor list.
Games
Sequential - reveals an
entire
image on each
step.
6x1 is a 6-square
grid
especially suited for race games.
3x2 and 3x3 are displays in
which
pictures
are revealed from left to right in rows.
EEG AVI Screen allows AVI files to be played one frame
at a time. A sample AVI file is
provided with the software to allow vendors to view the software functionality for the purpose of developing
and marketing games.
Selecting Game Images
Each screen
type is
designed to be used with a specific
group of game pictures.
All the
pictures
that are used for game feedback are in the
GamePics folder.
The GamePics
folder
is already selected by default in
applications that ship with the J&J software.
In
the GamePics folder are sub-folders containing specific types
of pictures. You can select these
subfolders by clicking the Game Windows button
.
All of the subfolders with names, such as
animals, travel, etc.
are scenic images designed to be used with screens that reveal images
randomly.
If you
purchase
games developed by other vendors for use
with J&J applications you will need to click the Game Windows
button and
select a game picture folder provided by your vendor.
The game picture folder you choose should have the same name as
the name of the Screen you select.
EEG Bar Displays, Peak Graphs,
Peak History Graphs
Bar
Displays show peak
microvolt
values of
individual bands. All bar displays
assume that band 1 is reinforcement, band 2 is inhibit and band 3 is
artifact
inhibit.
There are two
types
of bar displays: with threshold and
without threshold. Displays with THR in
the name indicate that the threshold is visible on the bar. The threshold indicator line reflects the
value of the threshold setting on the band associated with that bar. In THR displays, the bar changes color when
it crosses the threshold.
Peak Graphs are line graphs
showing
all of
the bands in peak microvolts.
Peak
History Graphs
show live
statistics of performance. They
show overall success rate for the
current task (the steps or reinforcements per minute called “Score
Now”), the overall score for the previous
task (“Score Last”),
and the peak microvolt averages for each band with markers to show task
boundaries.
Return to Index
Recording EEG Data
To record EEG
data
on a continuous basis simply click the
Start Task button at the bottom of your screen. If
you instead want to break the recording intervals into short
segments that are sequenced automatically, first click on the Select
Task
button
, select
the desired task(s), then click the Run Selected Task button. When you
are
ready to begin running your selected task(s), click Start Task. The number to the left of Start/Running Task
indicates the number of seconds the task will run unless you change it. You can use the up/down arrows to the left
of this number to increase or decrease the length of the task. You can manually end the task at any time by
clicking End Task.
The Task
Averaging
button
opens a window that lists the mean
average for each signal and each task as it is recorded.
Pausing and Unpausing Feedback Screens
The Start
Task/End
Task buttons and Pause Task
and Unpause Task
buttons
have similar functionality. Both the End
Task and the Pause Task buttons
pause the data recording and the feedback screen. Both
the Start Task and the Unpause Task buttons reset the
feedback screen and restart recording.
But when Start Task is re-clicked, it increments to the next
task (if a
programmed task has been selected), while Unpause Task resumes
recording of the
same task.
The Pause
Game
button
does
not effect recording. It merely freezes
the feedback screen while
the underlying data recording continues.
In feedback
screens
only the Pause and Unpause Task buttons
and the Pause/Unpause Game button
are visible. This means that in dual
screen mode both sets of controls will always be available, but in
single
monitor mode you will need to back out of the feedback screen if you
want to
end a task.
Return to Index
EMG Displays
A “W”
specified before a screen name in the
Select Screen Bar indicates that a wide bandpass filter (20-400Hz) has
been
applied to EMG signals displayed in that screen. The wide bandpass
filter
eliminates artifact frequencies outside of the EMG range. If no “W”
appears before a screen name, an additional narrow band filter (100-400
Hz) has
been applied to EMG displays on that screen.
This narrow band filter eliminates heart artifact from the EMG
signal. Within the display, narrow band
EMG signals will be labeled with an “N”. For
example, FFT_EMG A N or EMG-B RMS N.
The A or B refers to the input receiving the
EMG signal. Wide band EMG signals are
labeled within displays with a “W” as follows: EMG-B W
or EMG-A
RMS W.
The signal is then passed
through a 50 or 60 Hz notch filter to eliminate power line artifact.
(For more
information please see “Eliminating Electrical Line Frequency Noise” in
the
“Configuring the Software” section.
Unless the screen is labeled
“Raw” or “FFT” it contains EMG RMS (Root Mean Square) signals. Most EMG biofeedback screens display EMG RMS
signals. EMG RMS screens are defaulted
to AutoOffset ON so that the Signal Up/Down buttons are disabled. This
is
because evaluating the signal in relation to zero is essential to EMG
RMS
measurements. Moving the signal up so
that the zero is no longer visible on the scale would not be desirable
for this
display.
W-EMG Displays
The following display
descriptions apply to both wide band and narrow band EMG displays. The screen choices are identical, but a
filter to eliminate heart artifact has been applied to the signals in
the
narrow band screens. Wide band displays
show a greater degree of muscle activity so normally you will want to
use wide
band displays unless the site you are monitoring is contaminated by
heart
muscle artifact. Narrow electrode
muscle placement and placement far from the heart will have less
artifact than
wide placement and electrode sites close to the heart. You can
determine if
heart artifact is a problem if there is an observable R-wave point in a
raw EMG
display and/or a rhythmic pulsing in the 100-400 Hz range of an EMG FFT
display. If heart artifact is present,
use the screens without the “W” designation (the narrow bandpass
screens).
Single Signal Displays
The following EMG displays
contain only one signal per line graph space:
Raw EMG
A line graph display with vertical
microvolt scale and horizontal time scale, which provides a very fast
response
display of raw EMG activity. Because
this display responds very fast to very small movements, it is useful
for EMG
activity that is fleeting or weak. In
multi-channel versions, the signals appear as parallel stacked line
graphs with
scales locked together so that the Signal Control Bar buttons modify
all
signals together. (Pictured top left in the “Check Signals” screen on
page 16.)
RMS EMG With No Threshold
A line graph display with a
single integrated EMG signal. In
multi-signal applications, each signal is displayed in its own graph
with
linked vertical scales in microvolts and one common horizontal scale in
time
units.
RMS EMG With Threshold
EMG THR Down and EMG
THR UP displays contain one signal and one adjustable threshold
line.
In multi-channel displays
for symmetrical muscle training, parallel line graphs, each
with an RMS
EMG signal and down threshold are displayed, with their amplitude and
time
scales linked together so that GainUp/Down and Faster/Slower Graph
buttons modify all signals together.
Thresholds may be adjusted
individually for each signal. Autogain and Autothreshold are defaulted
to ON
for this display, but you can turn this feature off and manually adjust
the
signals. Background and fill colors may vary in different applications.
In multi-channel displays
for non-symmetrical muscle training, parallel line graphs, each
with an
RMS EMG signal and down threshold signal are displayed as above so that
the
displays superficially look identical, but their amplitude and time
scales are
not linked so that all signal adjustments are performed independently.
FFT EMG
A two-dimensional fill graph
on showing the frequency distribution of the EMG signal using a Fast
Fourier
Transform calculation. The bottom scale
represents frequency in Hz and the vertical scale represents amplitude.
Useful
for showing immediate frequency changes.
The display on the left shows a normal FFT signal, while the one
on the
right shows how effective this display is for revealing electrical line
frequency signal contamination, which appears as regular 50 Hz or 60 Hz
spikes
within the display. (For more information, see “Artifact Detection” in
your
Hardware Guide.)

3D FFT EMG
Same as
above but shown in 3-D with a third axis representing
units of time. This display is
particularly useful for highlighting even small frequency variations
over time.
This display is processor intensive and should be avoided if you have a
slow
computer.
EMG Bar With Threshold
In this display, an EMG
signal is represented as a bar graph, with a vertical microvolt scale
and a
digital display of the current signal value at the bottom.
An adjustable threshold line bisects the
bar. In multi-channel displays, each
channel is represented by another parallel bar graph in a different
color.
Multiple Signal Displays
The following displays
contain multiple signals within a single line graph.
These displays do not contain thresholds since it would be
confusing to have multiple signals, each with its own threshold,
sharing the
same space. Signal, background and fill colors may vary in different
screens
and applications.
RMS EMG
A display of two EMG signals
within a single line graph, color-coded to differentiate between the
Input A
signal and the Input B signal.
RMS EMG With Differential Fill
A vivid dynamic movement
display of the difference in muscle activity of two different EMG
sites. The color
of the signal with greater muscle activity fills the area between the
two signals. Useful for
evaluating/training opposing muscles or for evaluating/training
symmetrical
relaxation of two muscles or groups. EMG
Bottom Fill is an EMG Differential Fill display similar to the
above except
that the fill area of the stronger signal extends to the bottom of the
display,
instead of just filling the area between the two signals.
EMG Pattern
Also referred to as “EMG
Training”. A display of two EMG signals
with an adjustable pattern template for muscle training.
In this display, the Modify Signals button
opens the Myo Parameter Box, which you can use to adjust the pattern
template
by entering values for Cycles per Minute, Valley-to-Peak Time,
Hold-on-Peak Time,
Peak-to-Valley Time, Hold-in-Valley Time.
EMG Display Features
The Signal Up, Signal Down
and AutoOffset buttons are disabled in all EMG screens.
This is because EMG RMS displays and
spectral displays are most useful with a fixed zero at the bottom of
the scale,
and Raw EMG displays have a fixed zero in the middle of the screen.
Smoothing the Signal
For display purposes you may
want to smooth the signal. The
underlying data recording will not be affected. In
the EMG displays, the Modify Settings button opens the EMG
Filter Dialog window. Click on the Modify
Settings button
and
click on the Down/Up arrows next to the
“EMG Filter” box or type a numerical value in the box. A good general
purpose
filtering range for EMG RMS signals is from 30-60.
Changing Sweep Speed
For all signals except FFT,
the Faster Graph
and
Slower Graph
buttons
are used to increase/decrease the
time span that is displayed on the screen.
For FFT displays the Faster Graph & Slower Graph buttons
change the
frequency span of the FFT displays. It
is recommended that you leave FFT displays at the maximum frequency
scale.
Setting Thresholds
Please see the “Adjusting
Thresholds” section.
Return to Index
ECG, HR, and HRV Displays
Individual physiology may
vary and some individuals may have an indistinct R-wave point in their
ECG
signal. The software needs to be able to detect the R-wave in order to
calculate the IBI (Inter-Beat Interval) for HRV displays. If heart rate
tracking appears to be unreliable, try a right ankle, left wrist
placement with
either gel-free or disposable electrodes instead of wrist-to-wrist
placement.
RAW ECG
Pictured to the left below,
Raw ECG is a broadband signal display of electrical activity between
the two
electrodes, which includes the ECG waveform plus EMG activity at the
electrode
sites. The only purpose of this signal is to help you evaluate the
quality of
the ECG signal and the amount of artifact present in the signal. This display is not used for feedback. If
you get an unexpected reading on an HRV display, you can use the raw
ECG signal
to evaluate whether the anomaly is due to EMG artifact, indistinct
R-wave, or
missing or premature beats.
ECG FFT
Not intended for feedback,
this display is present on the Check Signals screen of HRV Breathing
applications
in order to evaluate electrical line frequency artifact, which appears
as
regular spikes at 50 or 60 Hz as shown below along with a contaminated
Raw ECG
signal display to the left. Please compare these with the
uncontaminated Raw
ECG and ECG FFT displays pictured above. (For more information on
signal
artifact, see “Checking For Good Signal Data” in the “Overview of the
Software
Interface” section and “Artifact Detection” in your Hardware Guide)
HR
The Heart Rate signal
reflects beats per minute and is recalculated and updated in the
display on
each beat. No averaging is performed.
This produces a heart rate waveform display which is a good
graphical
representation of HRV (Heart Rate Variability). This
signal is used in line graph displays along with the
Respiration waveform so that the impact of correct breathing on HRV can
be
easily observed as (1.) increases in variability in the HR waveform and
in (2.)
increased synchrony between the HR and Respiration waveforms. (Examples are pictured in the HR+Resp1+Resp2
display in the “Overview of Screens and Displays” and under “Breathing
Pattern”
in the discussion “Respiration Displays” above) The
Breathing Pattern screen has a unique way of displaying HR
that differs from the standard graph display. When a new heartbeat
occurs, its
value in beats per minute is plotted as a square dot and a line is
drawn from
each previous beat to each new beat, producing a graphing point
display.
The Heart Rate signal is
also found in bar displays (Pictured in the second Check
Signals screen
example in the “Overview of Screens and Displays” section) and in digital
displays (Pictured under “Numerical Displays” above).
HRV_30
This is a display of heart
rate variability averaged over 30 seconds.
The software measures the minimum and maximum heart rate values
for 30
seconds, then averages the difference of these values and displays them
in a
line graph. HRV_30 is a good graphical display of increasing or
decreasing HRV
trends.
DFT
The
DFT (Discrete Frequency Transform) display is a spectral display
for HR
with a horizontal scale of 0 -.4 Hz.
All the peaks at different frequencies in this display represent
the
power of different rhythms present in the inter-beat interval (IBI)
measurement. The calculation always
includes a fixed number of beats, either 64 or 128 beats depending upon
the
application, recalculated as each new IBI measurement is received. For example, in a subject with a breathing
rate of 6 breaths per minute (equaling one breath per 10 seconds = .1 cycles per second = .1 Hz) a cyclic
variation of heart rate every ten seconds would occur.
This would result in a peak in the DFT
display in the green area at .1 Hz on the scale.
The bands of color in the
DFT display represent the frequency ranges of the Very Low Frequency
(VLF), Low
Frequency (LF), and High Frequency (HF) signals as follows:
VLF= 0 - .05 Hz
LF = .05 - .15 Hz
HF = .15 - .4 Hz
The total power of all
frequencies is related to autonomic balance ( for example, people
suffering
from depression tend to have lower total power). VLF
is thought to be influenced primarily by the Sympathetic
Nervous System (SNS) with slight Parasympathetic, or Vagal, influence.
The VLF
value rises when people are intellectually engaged in activities such
as thinking,
worrying, or self-criticizing. LF seems
to be related to both SNS and Vagal activity associated with
barorecepter
regulation. The goal of breathing training is to achieve a peak in the
.08 to
.12 range of the LF band, allowing for variation in the breathing rate
of the
individual. This peak indicates
increased Vagal tone. Respiration is a
major influence over HF activity. Decreasing the breath rate causes
this value
to go down, while values of 10-24 breaths per minute cause the HF value
to go
up.
VLF,LF,HF
The values for each of these signals,
which represent
the summation of power in a particular band,
are displayed at the top of each DFT display (shown above) and
are also
used in line graph displays (shown below), bar graph displays (left)and
digital
number displays.
Respiration Displays
Respiration Line Graph
The respiration waveform
displayed in a line graph has a 0-1000 scale used to depict the
relative
stretch of the respiration sensor belt. This signal may be displayed
within in the
following display formats:
1.
A single
respiration
waveform may appear in a single line graph
2.
Multiple
respiration
signals may appear in a single line graph (See
HR+Resp1+Resp2 display pictured and described in the “Overview
of
Screens and Displays” section above)
3.
Multiple
respiration
signals may appear separately in stacked line graphs (See Resp1+Resp2
display
pictured within the second example Check Signals Screen and described
in the
“Overview of Screens and Displays” section above)
4.
A
respiration signal
may appear in combination with other types of signals as in the
Breathing
Pattern display below.
For those who prefer to do
breathing training without a pacer, a respiration
display
is a good choice.
Return to Index
Breathing Pattern
Three different signals are
displayed on this screen – a blue/orange breathing pacer, an aqua respiration wave form, and a red heart rate
variability wave form.
The effect of correct
breathing can be easily observed as an increase in heart rate
variability and
an increase in synchrony between the HR waveform and the Respiration
signal. The HR signal can be hidden
with the Show/Hide button if just the pacer and Respiration signal are
desired.
Adjusting the Breathing Pacer
To adjust the breathing pacer, click on
the small
green box near the top of your screen next to the word “Pattern”. The white dot that appears indicates that
the breathing pattern signal has been selected. The
buttons in the left hand signal control bar are used to
modify all signals on all screens. By
selecting “Pattern” you have assigned the signal control bar to the
breathing
pattern signal so that you can modify it.
Click on the Modify Settings button
In the Breath Parameter box,
adjust the target breath rate and inhalation time.
These are expressed as percentages of the total breath time,
with
the sum of inhale time, inhale hold, exhale time and exhale hold equal
to 100%. By adjusting the inhale time the
exhale time
automatically adjusts, and adjusting the exhale time automatically
alters the
inhale time. Also notice the changes in
the wave form. Click on the small x in
the upper right hand corner of the Breath Parameter window to close it.
BPM Bar Graph
Physiolab detects the completed
respiration cycle, re-calculates the BPM (Breaths Per Minute) each
cycle and
displays the result as a vertical bar graph.
BPM Numerical Display
Physiolab includes BPM as a digital
value in Numerical Displays on screens where is may be of value but is
not the
main signal of interest.
Return to Index
Skin Conductance and Skin Resistance
Displays
Skin
conductance is more commonly used to record data and display baseline
information. Skin Resistance is a more
sensitive measure of small changes in arousal which enables more
effective
visual displays. In the Sensor Test screen, Skin
Resistance is
used to test signal quality because it is a more sensitive measurement
and
because it is the signal directly measured by the instrument. Skin Conductance is a mathematically derived
value.
The scale
for all Skin Resistance signals is in KOhms.
For Skin Conductance displays the scale is in MicroMhos.
In applications with
EMG/ECG/EEG impedance testing, bouncing of the SR (Skin Resistance) Bar
Graph
is due to interference from the impedance testing function. This does
not
effect your ability to evaluate the signal quality based upon its bar
graph
color and this interference is not present in any of the feedback
screens
(because impedance testing is only a function of the Sensor Test
Screen). In applications without
EMG/ECG/EEG signals,
no impedance testing takes place. For more
information about the Sensor Test Screen, please see “Ensuring a Good
Connection”
in the “Overview of the Software Interface” Section.
The Check Signals screen
displays all signals available in the application and currently being
received.
As in the Sensor Test screen, the Skin Resistance signal is also used
in this
screen. For more information about the Check Signals screen, please see
“Checking for Good Signal Data” on page 5.
SR Graph
A line graph display with
the Skin Resistance signal scrolling from left. In multiple signal
applications, the SR Graph display is a stacked line graph, with one
graph for
each signal, with a common horizontal time scale and vertical scales in
KOhms
locked so that all signals are adjusted together.
Dual SR Graph
A single line graph with two
Skin Resistance signals sharing the space. Signal controls adjust both
signals
together.
SR Bar
A vertical bar display with
a Skin Resistance signal and an adjustable threshold line.
If the SR signal is too sensitive (large
swings up and down), click AutoGain OFF and click the Gain Down button
repeatedly until an acceptable range is achieved.
SC and SR Scroll
Scrolling chart displays of Skin
Conductance are
available combined with chart displays of Temperature or Skin
Resistance
values.
SCR Scroll
SCR (Skin Conductance Response) is
the degree of
change in the Skin
Conductance signal measured in MicroMhos per second.
Currently implemented in the software as a stacked scrolling
chart display, along with charts of Skin Conductance values.
Return to Index
Temperature Displays
Temperature can be represented
in the following display formats discussed under “Display Types”: line
graph,
rainbow fill graph, scrolling chart, bar graph, scrolling history bar,
and
numerical display formats. Specific
examples are described below.
Fahrenheit-Centigrade Button
The toggle button
switches
the display units between Fahrenheit
and Centigrade. The setting you are
using when you exit a session is saved until you manually change it.
TEMP Graph
A line graph display with a
Temperature signal scrolling from left. In multiple signal
applications,
stacked line graphs, one for each signal, with a common horizontal time
scale
and vertical scales in degrees Fahrenheit locked so that all signals
are
adjusted together.
Dual TEMP Graph
A single line graph with two
Temperature signals sharing the space. Signal controls adjust both
signals
together.
Dual Temp Scroll
A stacked chart display with
a single temperature signal which scrolls to the left in each chart.
Locked vertical scales in degrees
Fahrenheit and a common horizontal time scale.
TEMP Bar
A vertical bar display with
an adjustable threshold line. Clicking
the Zoom button to fill an entire screen with this display provides a
simple
effective feedback screen. For
threshold instructions, please see “Adjusting Thresholds” on pg. 21.
TEMP Numerical
Display
A numerical
digital
value for Temperature in degrees
Fahrenheit may be displayed in screens for which Temperature is not the
modality of focus, but may be useful to monitor as an auxiliary
measurement.
For an example, see “Numerical Displays” in the Display Types” Section.
Return to Index
Audio Features
There
are two audio icons in the left control bar. The Sound ON/OFF button
toggles
the speakers on and off. IMPORTANT: This
button must be ON in order
for audio and video features to work.
The
Select Sound button
brings
up the Sound and Game Control
window.
This window is used to set
up three types of media feedback:
1.
the computer
music synthesizer which generates
audio pitches that change with the signal,
2.
pre-recorded
audio files of spoken
words or sound effects that are triggered
when signal thresholds are crossed,
3.
pausing
and unpausing Windows Media player
audio-visual displays when thresholds are crossed.
You may have up to three
signals with audio feedback simultaneously functioning, each with a
different
type of media feedback. It is not possible to simultaneously run the
same TYPE
of media feedback on more than one signal in a display.
For example, you may have one signal with
audio file feedback, one signal with synthesizer feedback and one
signal with
Media Player feedback, BUT NOT three signals all with simultaneous
Media Player
feedback. IMPORTANT: If you set up
Media Player for more than one signal in the Sound and Game
Control
window, Media Player will not work.
The top box displays the list of signals
for which
audio is available in the current display screen. To
enable sound for a particular signal, click on the empty check
box next to the signal name. A checkmark will appear to indicate that
sound is
enabled for that signal. To disable sound for a particular signal,
click on the
check box to remove the checkmark next to that signal name. A checkmark simply turns the sound on or
off; it does not affect signal settings.
Audio feedback must be set
up separately for each signal. To modify
settings for a particular signal,
click on that signal name to highlight it.
Click on the left-hand Event Control drop-down box.
Select ABOVE or BELOW threshold depending
upon whether you wish audio feedback to sound when the signal crosses
above or
below the threshold line. (NOTE: Some
signals, such as EEG bands or a combination of skin resistance and
temperature,
have multiple thresholds yoked together so that together they are
considered to
be a single logical threshold. When all thresholds are simultaneously
satisfied, audio-visual feedback is triggered.
For these displays you will select the name of the yoked
threshold
instead of ABOVE or BELOW.)
After selecting your
threshold choice, click on one of the following three buttons: Voice
and
Music Files, Music Synthesizer or Media Player. For
further
instructions go to the section below which matches the button you have
selected. (To change the signal threshold see the instructions for the
Threshold button in the previous Modifying Signals section.)
Setting Up Music Synthesizer Audio Feedback
The Sound drop-down
list displays all available built-in musical instrument sounds. These
have a
pitch scale similar to a piano keyboard.
The numbers in the Lowest/Highest
Pitch boxes represent notes on the musical scale.
The usable range is approximately 20 for the lowest pitch to 100
for the highest pitch. The most audible and pleasant range will vary
for each
instrument. Each instrument has its own preset values, but you may wish
to
experiment and modify these to find a desirable range for each
instrument.
The current note played
always represents the difference between two signal values - the current signal value and one of three
reference values listed below:
(1) the signal value at the
point when you clicked “Adjust Reference”, or
(2) the scale value at the
bottom
of the display, or
(3) the threshold value.
You can anchor the scale to
either the “Fixed Signal Value”, the “Edge of Display” or the
“Threshold Value”
by clicking the small circle in front of your choice.
A dot will indicate your selection.
The Set Reference Pitch
slider changes the value of the reference note (displayed in the small
box to
the right). The slider is only moveable between the values specified in
the
Lowest Pitch and Highest Pitch boxes.
The reference pitch is the note that will be played when the
current
signal value equals the selected reference value. Musical
pitches will go up or down from this reference pitch to
the highest or lowest values specified in the Lowest Pitch and Highest
Pitch
boxes.
The Fixed Signal Value
is the signal value at the point when you click the “Adjust Reference”
button.
If Fixed Signal Value is selected, the reference pitch note will be
re-set to
the current signal value whenever you re-click the Adjust Reference
button. The
number in the box to the left of the Adjust Reference button is the
value of
the reference pitch note. The number in
the box to the right is the value of the current note.
If Edge of Display is
selected, the reference pitch is anchored to the scale value at the
bottom edge
of the display, with musical pitches becoming higher from that point.
If Threshold Value is
selected, the reference pitch (set with the slider) is anchored to the
threshold so that the musical pitch goes up as the signal rises above
the
threshold and down as the signal falls below the threshold (unless
Invert Pitch
is selected).
Normally, if the signal goes
up, the musical pitch goes up. Checking Invert Pitch causes the
opposite: the musical pitch becomes higher as the signal values become
lower
and the musical pitch goes down as signal values go up.
With Invert Pitch selected the reference
pitch should be set to a higher number.
The Signal Change Per
Note Step value determines the number of units of change on the
signal
scale required to produce the next pitch.
The Delay Next Note
value determines the time in seconds that must pass before the next
note is
played.
The length of the note is
dependent upon the musical instrument selected and is a fixed value
that cannot
be changed.
If Repeat Note is
checked, the same pitch will be repeated when there is no signal change
greater
than the Signal Change Per Note Step interval.
If unchecked, the signal must change the amount specified in the
Signal
Change Per Note Step box before the next note will be played.
If Don’t Skip Note is
selected, each pitch in the musical scale will be played with the time
interval
delay specified in the “Delay Next Note” box between each note. This provides a gradual, musically pleasing
transition between pitches, but can cause the pitch to lag behind the
signal
during sudden large signal changes.
Setting Up Pre-Recorded Audio File Feedback
USE3 comes with a wide selection of WAV
files with
pre-recorded voice feedback and sound effects.
A WAV file is triggered when a threshold is crossed. The entire
1-3
second file is played. If threshold is
maintained, then next WAV file will play after your chosen delay time.
To
set up audio file feedback, first enable (check) the signal you wish to
set up,
click on the signal name to highlight it, and make your threshold
choice in the
drop down box. Then click the Voice
and
Music Files button.
Click the Folder
button above. Double-click the Sounds
folder. Click on a sound collection
folder to select it, then click OK.
You may choose to play all of the files in
the folder
in order, play all of the files randomly, or you may choose to play
just one
file from the folder. Click the small
circle in front of “Play All in Order”, “Play All Random” or “Play
Selected” to
indicate your choice.
If you choose “Play
Selected”, you must click the drop-down box under Select Voice or Music
Files
and click on a specific file name.
Setting Up Windows Media Player Feedback
To
set up Media Player feedback, first enable (check) the signal you wish
to set
up, click on the signal name to highlight it, and make your threshold
choice in
the drop down box. Then click the Media
Player button and close the window.
If you are running single display
mode, Media Player will be
minimized in the task bar. It may not be visible
until the connected signal crosses the threshold. Click on Media Player
in the
task bar to bring up the Media Player window.
If you are running dual display mode, next click the
Select
Second Screen Icon, then highlight media player in the Screen list and
click
Apply.
Next
click Open at the top of the Media Player window. The
top drop-down box in the “Open” window
defaults to the Physiolab Media Player folder.
Click on a file name to select it , then
click Open. You can add your own audio
and video files to the C:\Program Files\J&J
Engineering\Physiolab\Media
Player folder or you can use the top drop-down arrow box in the Open
window to
browse to the location of any audio or video file that you wish to play
either
on your hard drive or on a CD in your CD-ROM drive.
Music files are normally stored in the folder My Documents\My
Music. In the “Open” window drop-down
box you may have to go up several folder levels to get to your CD-ROM
drive or
music file folders.
If
you are playing a music file in Media Player, the background
visualization will
default to the settings from the last time you used Media Player.
Media Player Troubleshooting
Changes
to the Select Sound window in Physiolab may deselect the play file in
Media
Player. Click Open in the Media Player
window and re-select the audio or video
file.
If your Media Player window does
not appear to be functioning
(clicking on Open does not work to select a file OR
the screen appears frozen while a file is playing), click on MBP Media
Player
in the bottom task bar to make sure Media Player is selected as an
active
window.
If Media Player is running, you
are getting no sound, and the Media
Player window says “Paused”
in the
lower left-hand corner, then check the following:
- Make sure the Sound ON/OFF
button
is pressed.
- Open the Select Sounds window with
the
Select Sound button
. Make sure
that
Media Player is listed as the
sound type for the signal you are expecting. If not, see the section
above on setting up Windows Media Player feedback.
If Media Player is running, you
are getting no sound, and the Media
Player window says “Playing”
in the
lower-left hand corner, then check the following:
- Windows Speaker volume controls in
Start>Control Panel>Sounds
- External Speaker connections to the
computer and to the power outlet
MBP
Player Error Message: Go to the Microsoft website and get the Media
Player
version 9 update. You need to have Windows98 Second Edition with all
Microsoft
update downloads or a later Windows operating system (WindowsXP
recommended).